Parameter Set
|
HSQCETGP |
Pulse Sequence |
hsqcetgp |
| Experiment Description |
-
- Heteronuclear Single Quantum Coherence Spectroscopy with gradient selection and INEPT enhancements.
- Correlates chemical shifts of 1H nuclei directly bonded to 13C sites
|
Scans per Slice
(ns) |
Dummy Scans
(ds) |
Recycle
Delay
(d1) |
FID Size
(TD, F2) |
# Slices
(TD, F1) |
F2/F1
Spectral
Width
(sw,ppm) |
Spectral
Center
(o1p/o2p, ppm) |
| 1x ns (2) |
16 |
2 |
1024 |
256 |
13/165 |
6.18/75.0 |
| Experiment Setup |
-
- Use getprosol or prosol button to read in proper pulse parameters
- Set cnst2 = 1JCH coupling constant: default cnst2 = 145 Hz
|
| Processing Notes |
-
- Fourier transform with the command xfb
- Manually phase F2 and F1 dimensions
- Use commands abs2 and abs1 for baseline correction
- For improved resolution, can try applying linear prediction in the F1 dimension. In the ProcPars tab set the following parameters:
- ME_mod = LPfr
- NCOEF = set to 3 times the number of cross peaks in your spectrum.
- LPBIN = 2*TD(in F1)
- Fourier transform the spectrum again with xfb
|